BlueLine module options
Module option
EMISSION-FREE STERILIZATION
Emission-free sterilization of hazardous substances with exhaust air filtration and condensate sterilization
Whenever a laboratory works with genetically modified organisms, infectious material or pathogens, the steam sterilizers used are subject to special technical requirements. These always apply in safety level 2 laboratories and in S1 laboratories whenever RG1 biological agents with sensitizing or toxic effects are to be autoclaved. In order to effectively deactivate contaminated waste, a fractionated pre-vacuum can be used for solid waste. In addition, the resulting condensate must be sterilized, which must remain in the autoclave for the entire sterilization process. Exhaust air filtration ensures the aerosol protection required from safety level 2.
This makes our FA module suitable for the sterile disposal of potentially infectious laboratory waste of protection classes 2 to 4 as well as for the sterile disposal of GMOs from laboratories of safety levels 2 to 4. Short pipe runs enable a constant sterilization temperature even in the exhaust air filter, which can be replaced particularly quickly and easily.
FA module exhaust air filtration with condensate sterilization
Our module option for the safe sterilization of hazardous substances enables efficient autoclaving with the highest safety standards: Aerosols in the exhaust air are already filtered out during the heating phase so that the room air remains uncontaminated. Chamber contents, the resulting condensate and the filter housing including filter are sterilized with each sterilization process. After successful sterilization, the condensate is disposed of via a dead space-free drain valve in the chamber floor, which enables particularly simple and contamination-free disposal of the condensate. In the event of a fault in the media supply or an unexpected program interruption, the FA module switches off automatically. The FA module exhaust air filtration with condensate sterilization can be optimally combined with VARIOKLAV® cooling, vacuum and drying modules.
Technical data
FA module for exhaust air filtration with condensate sterilization
| Filter housing | Pressure-resistant filter housing made of stainless steel |
| Pneumatically controlled valves, pressure and vacuum-tight | Safely isolate the disposal area in the event of a compressed air supply failure |
| PT-100 sensor measures filter temperature in filter housing | Sterilization time is only started when the set temperature is reached in the filter |
| Exhaust air filtration | Membrane filter μ = 0.2 μm, 0.2 m² hydrophobic |